CPSA is a preprocessing method used to remove interferences from data prior to building a regression model. The CPSA filter includes one or more of the following corrections:

  1. multivariate filtering of components defined by a set of clutter or "correction" spectra.
  2. detrending (aka baselining) of the input data in windows of variables.
  3. an integrated thickness/pathlength correction.

CPSA Panel

The CPSA tools are accessed through the CPSA panel of Analysis GUI (click on the "CPSA Settings" button just below the top status panel). The CPSA settings panel is only available when using the PCR or PLS Analysis methods and only enabled when calibration data has already been loaded. The default settings for the panel are shown below.

Cpsa panel empty.png


Four different options categories are available on the panel:

  • Correction Spectra - Allows selecting correction spectra which describe background components which should be removed from the spectra prior to analysis.
  • Regions/Baselines - Defines regions of the spectra and the baseline order to apply to each region.
  • Thickness Correction - Determines whether or not to use thickness/pathlength correction (requires thickness/pathlength reference in calibration data)

Correction Spectra

The list on the left of this option shows the file(s) which have been selected by the user as correction spectra (aka background or clutter spectra) which should be removed from all calibration and test spectra. To perform the filter, the correction spectra are first decomposed using a Principal Component Analysis (PCA) model containing a user-specified number of principal components or "factors". The resulting factors are then used as the filter.

If no correction spectra are loaded, no background removal will occur, but all other filter steps (see other options) will still be completed.

Adding Correction Spectra

To add correction spectra to the filter, click on the Add button and select the file type to read (e.g. "Hamilton Sundstrand ASF files") and click OK.

Browse to the folder containing the file(s) to load, select one or more files (using shift-click or control-click to select multiple files), and click "Open".

The files will be loaded and added to the list. You will be advised if the spectral axis of the loaded files appears to be incompatible with the currently loaded x-block data. Any incompatible spectra will not be loaded.

Cpsa correction loaded.png

Removing Correction Spectra

Any file that has been loaded as a correction spectrum can be removed from the filter by simply clicking on the filename in the list on the left, and then clicking on the Remove button. The file will be discarded. Note that after discarding any spectra from the filter, any currently-loaded model will be cleared and have to be recalculated because the filter has now changed and the model is no longer current.

Choosing Number of Factors

The CPSA filter calculates a PCA model from the correction spectra and keeps some number of factors. The edit box on the right of the control specifies the number of factors to use in the filter. The higher the number, the more degrees of freedom the filter will remove (and the increased chance of removing useful signal). In general, a lower number of factors is recommended unless the filter is not performing sufficiently for a given application.

Viewing the Correction Spectra

The selected correction spectra can be viewed in a plot by clicking the Plot button. The spectra selected on the left will be displayed, but all loaded correction spectra can be viewed using the Plot Controls which will appear along with the plot of the spectra. Regions of these spectra can be excluded using the Select button and Edit/Exclude or Include menu options in the Plot Controls.

Regions/Baselines

By default, all spectral regions marked as "included" in the X-block data will be used in the calibration model. However, the Regions/Baselines option of the CPSA panel allows selection of specific spectral sub-regions to include (and, conversely, to discard other regions). In addition, the selected sub-regions can be detrended (i.e. baselined across the entire selected region).

To select one or more specific sub-regions, click the Edit button next to the Regions/Baselines option on the CPSA panel. This will bring up the Regions and Baselines GUI. Initially, no regions will be selected and you will see the mean of all the calibration spectra and an empty table at the bottom of the GUI (see figure below.) Note that all changes within the Regions and Baselines GUI take effect immediately.

If you want to view a specific spectrum or spectra in the calibration data instead of the mean, select the Plot/Rows menu item in the Plot Controls, then select the spectrum/spectra to view in the Plot Controls.

Cpsa regions empty.png

The toolbar on this figure provides several options:

Cpsa regions toolbar.png

  • Folder Icon - import regions and baselines from a CPSA32 .cri file
  • [+] Icon - add a new region (click and drag)
  • [-] Icon - remove the region currently selected in the table at the bottom of the GUI
  • Double-arrow Icon - reverse the x-axis direction
  • Check Mark Icon - accept the current regions and close the GUI

Adding Regions and Baselines

To begin selecting regions, either enter the starting and ending frequencies in an empty line in the table at the bottom of the GUI or click the [+] toolbar button and click-and-drag a new region on the interface. In the figure below, the user has manually entered a starting and ending value creating the first region (shown as shaded blue region in the plot.)

Cpsa regions first.png

Once created, regions can be edited by either directly editing the values in the table, or by clicking and dragging the dark region markers on the left and right edges of each region in the plot.

In addition to specifying the included spectral range, the table also allows specifying the order of polynomial to use as the baseline. Note that although the terminology here is "baseline", the calculation is, in fact, a "detrend" in which the polynomial is fit to the entire range of windowed intensities.

In the example above, the specified window will be baselined using a 2nd order (Quadratic) polynomial.

Removing Regions

To remove a region, either select the corresponding row in the table at the bottom of the GUI and press the [-] toolbar button, or right-click the region and select Remove Region from the context menu. In the example below, the blue region has been right-clicked giving the Remove Region context menu.

Cpsa regions delete.png

Thickness Correction

The thickness / pathlength correction option allows automatic detection and correction of pathlength errors. To make this correction, the calibration data must be loaded with pathlength values in the sample axisscale of the X-block (note that the name of the axisscale must be "pathlength"). The values supplied will be used to calculate a correction model which, when applied to future data, will attempt to correct any pathlength variations detected in that data.

To enable this feature, check the Use Thickness Correction checkbox. If no pathlength data was supplied with the calibration X-block spectra, you will be prompted to specify a default value to assign to all spectra. These default values, or any values supplied when importing the spectra, can be modified by clicking the Edit button to the right of this option.

Editing the pathlength will open a DataSet editor for the X-block with the Row Labels tab selected. The Axis Scale column of this interface contains the pathlength values. These values can be manually corrected, or copied and/or pasted from another application. To copy the current values or paste in new values, click on the Axis Scale button at the top of the interface, then select the menu: Edit/Copy or Edit/Paste. All changes in the DataSet editor take effect immediately.

Uncheck the Use Thickness Correction checkbox to disable the thickness correction feature.

Force PC80 Compatible Preprocessing

If this option is unchecked, the user may use any preprocessing available in the custom preprocesing interface. The model created will be compatible with Eigenvector Research software such as Solo_Predictor, Solo, and PLS_Toolbox but may not be compatible with the Hamilton Sundstrand PC80 and CPSA32 applications.

When checked, the preprocessing selected in the pull-down menu to the right of the option (including: None, Mean Centering, Autoscaling) will be used and any other preprocessing manually selected from the Custom preprocessing GUI or the preprocessing menus will be removed. This mode assures that the model created will be as compatible as possible with the PC80 and CPSA32 applications.